Infinite® M200 PRO and F200 PRO NanoQuant - Typical performance values

Light Source UV Xenon flashlamp
Wavelength selection
Infinite M200 PRO
Quad-4 Monochromators system
(2 excitation monochromators)
Bandwidth Ex: < 9 nm for λ> 315 nm; < 5 nm for λ≤ 315 nm
Wavelength accuracy < ± 0,5 nm for λ> 315 nm; < ± 0.3 nm for λ≤ 315 nm
Wavelength reproducibility < ± 0,5 nm for λ> 315 nm; < ± 0.3 nm for λ≤ 315 nm
Infinite F200 PRO NanoQuant
 
Bandwidth 260 nm (5 nm bandwidth)
280 nm (3 nm bandwidth)
   
Wavelength range 230 - 1000 nm
Absorbance
Detection limit dsDNA concentration 1 ng/µl
Reproducibility of one sample (CV)
(50 ng/µl)
< 1%
Uniformity (50 µg/ml) < 3%
Ratio 260 / 280 nm (50 µg/ml) ± 0,07
Precision @ 260 nm < 0,2 %
Accuracy @ 260 nm < 0,5 %
Measurement range 0 - 4 OD
   
Detectors UV Silicon photodiode
   
Plate formats 6- to 384- well plates, cuvettes, NanoQuant Plate™
   
Temperature control Ambient +5°C up to 42 °C
   