Sunrise™ microplate reader - Specifications

Sunrise general specifications
Optical system 12 measurement channels, 1 reference channel
Filters up to 6
Wavelength range 340 to 750 nm
Bandwidth @ 50% transmission 10 +/– 2 nm
Measurement range 340 to 399 nm: 0 to 3.0 OD, 400 to 750 nm: 0 to 4.0 OD
Measurement time Single wavelength: 6 seconds
Dual wavelength: 8 seconds
Kinetic repeat time: 5 seconds
Accuracy
0.0 to 2.0 OD/492 nm < (1.0% + 0.010 OD)**
2.0 to 3.0 OD/492 nm < (1.5% + 0.010 OD)**
Linearity
340 to 399 nm: 0.0 to 2.0 OD < 2%, R2 >= 0.999
400 to 750 nm: 0.0 to 2.0 OD < 1%, R2 >= 0.999
2.0 to 3.0 OD < 1.5%, R2 >= 0.999
Precision
0.0 to 2.0 OD/492 nm: < (0.5% + 0.005 OD)**
2.0 to 3.0 OD/492 nm: < (1.0% + 0.005 OD)**
Resolution 0.001 OD
Interface RS 232 C (remote control)/parallel
Shaking Linear shaking in 4 different modes
Physical data Width: 28.5 cm
Depth: 34.0 cm
Height: 14.5 cm
Weight: 6.2 kg (8.6 kg incl. all options)
Specification tunable wavelength option
Wavelength range 400 to 700 nm using gradient filter/340 to 750 nm using standard filter
Accuracy 0.0 to 2.0 OD/492 nm < (1.5% + 0.010 OD)**
Linearity 0.0 to 2.5 OD/492 nm < 2%, R2 >= 0.999
Precision 0.0 to 2.5 OD/492 nm < (1% + 0.005 OD)**
Specification temperature control option
Temp. range RT to 42 °C (resolution 0.1 °C)
Temp. accuracy typical +/-0.2 °C (max. +/-0.5 °C)

Specifications are subject to change without notice
** better than x % of measurement value plus corresponding OD value.
Note: All deviations from the measurement value are meant in positive and negative direction.