General  
Number of disp. channels 1–4
Wash head types 96HT for high speed washing in 96-well format
384HT for high speed washing in 384-well format
96 indexing for flexible washing in 96- & 384-well formats
Wash volume, 96-well format 50–3000 μl in increments of 50 μl
Wash volume, 384-well format 10–1000 μl in increments of 10 μl
   
Performance  
Wash times, 96-well format 15 secs for 1-cycle wash incl. crosswise asp. step
(disp. vol: 300 μl/well, disp. rate: 5, asp. rate: 5, 96HT head)
Wash times, 384-well format 15 secs for 1-cycle wash incl. asp. step
(disp. vol: 100 μl/well, disp. rate: 5, asp. rate: 5, 384HT head)
Dispense accuracy @ 300 μl, 96-well format ≤ 5 % across plate, gravimetric measurement*
Dispense accuracy @ 100 μl, 384-well format ≤ 5 % across plate, gravimetric measurement**
Dispense uniformity @ 300 μl, 96-well format ≤ 3 % across plate*
Dispense uniformity @ 100 μl, 384-well format ≤ 4 % across plate**
Residual volume, 96-well format ≤ 2 μl per well, crosswise asp.*
Residual volume, 384-well format ≤ 2μl per well, single asp.**
   
Vacuum range for vacuum filtration -50 to -300 mBar
   
Power  
Power supply 115 V / 60 Hz
230 V / 50 Hz
Consumption < 300 VA
   
Physical  
External dimensions (height/width/depth) 28.8 cm (11.3 inch) / 38.8 cm (15.3 inch) / 43.0 cm (16.9 inch)
Weight 15 kg (33 lbs)
   

* disp. rate: 5, respectively asp. rate: 5, 96HT head, wash buffer plus 0.1% Tween, Greiner® flat-bottom plate